Voyage: IN2021_V02       [details]
Subject: Serial output was not connected
When: 2021-04-20 06:19:00 UTC

- entered into Elog/Everlog as 2021-04-20 06:19:00
Log: in2021_v02/Atmospherics
Equipment: CCNC

Cloud Condensation Nuclei Counter [view]
Position: Latitude: -46.9717 Longitude: 142.3815 Position from underway logging
Action: Fault
Encoding:HTML
Comments:

Serial output was not connected, preventing GORE from collecting data.

The serial configuration was also screwed up, preventing the strings from having any sensible data.

Trevor plugged in the serial cable, I have fixed the serial config and restarted the software. all appears to be working now.