Please login if you have access to particular applications.
Event details
Voyage: | IN2020_V06 [details] |
Subject: | Clean and signal check |
When: | 2020-10-19 06:34:39 UTC - entered into Elog/Everlog as 2020-10-19 06:34:39 |
Log: | in2020_v06 |
Equipment: | Fluorometer
Fluorometer [view] |
Position: |
Latitude: -55.21167 Longitude: 158.897
- entered into Elog as 55 12.7 S, 158 53.82 E |
Action: | Maintenance |
Encoding: | HTML |
Comments: | Fluorometer has had numerous occasions where output value is below calibrated darkcounts. Instrument was removed and rigorously cleaned in fresh water. Full scale was tested with test stick and worked. Reinstalled for monitoring. |