Event details
Voyage: | IN2019_E01 [details] |
Subject: | Flow cal |
When: | 2019-08-01 01:34:09 UTC - entered into Elog/Everlog as 2019-08-01 01:34:09 |
Log: | in2019_e01_Atmospherics |
Equipment: | CPC 3772
Condensation Particle Counter (TSI 3772) [view] |
Position: |
Latitude: -13.88717 Longitude: 146.06583
- entered into Elog as 13 53.23 S, 146 03.95 E |
Action: | Maintenance |
Encoding: | HTML |
When doing the flow cal with lab air: | |
Comments: | Set the CPC back up and reconfigured the plumbing so that the orifice pressure on the CPC will register OK. Connected up to the strong vacuum so that the orifice pressure reads OK (this is fine because the MFC is maintaining the flow actual flow). When doing the flow cal with lab air:
When doing the flow cal inline:
The inline set point is really variable, but after an average of 10 tests, we& 39;ve decided to leave it as above. |