Please login if you have access to particular applications.
Event details
Voyage: | IN2019_E01 [details] |
Subject: | Switched instrument off to packup for repair |
When: | 2019-08-01 00:26:44 UTC - entered into Elog/Everlog as 2019-08-01 00:26:44 |
Log: | in2019_e01_Atmospherics |
Equipment: | SMPS
nano Scanning Mobility Particle Sizer (Grimm) [view] |
Position: |
Latitude: -13.88767 Longitude: 146.06517
- entered into Elog as 13 53.26 S, 146 03.91 E |
Action: | Outage |
Encoding: | HTML |
Comments: | Error 32 suggests that the peltier cooler is broken and GRIMM are asking to send it back to them in Germany. Ian is going to try to fix it in Hobart before sending it back, so I& 39;m packing everything up, including the PC. |