Please login if you have access to particular applications.
Event details
Voyage: | IN2019_E01 [details] |
Subject: | Zero check started |
When: | 2019-07-22 08:59:06 UTC - entered into Elog/Everlog as 2019-07-22 08:59:06 |
Log: | in2019_e01_Atmospherics |
Equipment: | CPC 3772
Condensation Particle Counter (TSI 3772) [view] |
Position: | Position not recorded in Elog/Everlog. |
Action: | Maintenance |
Encoding: | HTML |
Comments: | Re-installed instrument after 12M service, positioned next to CCN inlet. Calibrated flows (gilibrator = 999ccm (n=5), Alicat set= 1.000LPM, Alicat read= 1.000LPM) HEPA zero check started 0900hrs UTC. |