Please login if you have access to particular applications.
Event details
Voyage: | IN2019_E01 [details] |
Subject: | Annual maintenance |
When: | 2019-07-21 15:00:00 UTC - entered into Elog/Everlog as 2019-07-21 15:00:00 |
Log: | in2019_e01_Atmospherics |
Equipment: | SMPS
nano Scanning Mobility Particle Sizer (Grimm) [view] |
Position: | Position not recorded in Elog/Everlog. |
Action: | Maintenance |
Encoding: | HTML |
SMPS maintenance including: | |
Comments: | SMPS maintenance including: - Cleaning the external impactors with kimwipe and alcohol - the nozzle on the impactor (not the preimpactor) may have been installed backwards. See new procedure for proper installation. - Cleaning the impactor at the top of the column with kimwipe and alcohol. - Reapplicaiton of a thin layer of grease to all the impactors - Cleaning of the DMA and all surfaces with kimwipes and alcohol. The DMA itself was very clean, and it was just the sections around the external screws that had a bit of corrosion on them. The brass connections looked reasonable, however one side (presumably the inlet) was dirty on the inside. - Changed the excess air filter with a new one. |