Event details
Voyage: | IN2015_C02 [details] | ||
Subject: | Restart logging | ||
When: | 2015-12-11 04:14:02 UTC - entered into Elog/Everlog as 2015-12-11 04:14:02 | ||
Log: | GSM_in2015_c02 | ||
Equipment: | EM122
Kongsberg EM122 12kHz multibeam echo sounder [view] | ||
Position: |
Latitude: -35.25283 Longitude: 130.72083
- entered into Elog as 35°15.17'S, 130°43.25'E | ||
Action: | Start | ||
| |||
Comments: | EM122 _ fault finding. Shut all systems down apart from pre-amp. Tested voltage on TP17/18 on each pre-amp card, all acceptable and within range as per manual. Checked seating of TX cards in EM122 TRU, all appear ok. Restart system, initial error that ping status is faulty as before however ack this error restart pinging and system work ok. She lives to ping another day, as to the root cause of this problem , atm, no fecking idea. |