Voyage: IN2016_V02       [details]
Subject: bad headiing - see log
When: 2016-03-19 04:50:41 UTC

- entered into Elog/Everlog as 2016-03-19 04:50:41
Log: GSM_in2016_v02
Equipment: EM122

Kongsberg EM122 12kHz multibeam echo sounder [view]
Position: Latitude: -46.8251 Longitude: 141.8662 Position from underway logging
Action: warning
Testing ended 10:06 UTC
Comments: strange did - restart and see what happensAfter 5 attempts booting the transceiver and finally once the em122 -2 pc, nothing worked.Can see outgoing pings in hydrophone using internal triggeringDecide to swap to the spare HWS. All BIST were OK except TX36 saying non-unique hardware.Found no improvement. Still saw pings - a little more infrequent.Switched back to em122-2 PCTesting ended 10:06 UTC