Please login if you have access to particular applications.
Event details
| Voyage: | IN2018_V05 [details] | ||
| Subject: | Unmatch between CTD Temp and Salinity Calibration | ||
| Event Identifier: | IN2018_V05_001 | ||
| When: | 2018-10-16 22:31:15 UTC - entered into Elog/Everlog as 2018-10-16 22:31:15 | ||
| Log: | in2018_v05 | ||
| Equipment: | CTD
CTD - Seabird 911 with 24 Bottle Rosette [view] | ||
| Station/Location: | Test station (Station no: 001) | ||
| Position: |
Latitude: -45.8615 Longitude: 148.70917
- entered into Elog as 45 51.69 S, 148 42.55 E | ||
| Action: | calibration | ||
| |||
| Comments: | There is an apparent difference - bias between the two CTD temperature sensor of ~0.001 C at a depth of ~2720 m, potentially due to calibration, it was decided not to replace temperature sensors. Salinity looks more stable ranging from 0 to 0.0005 ups. | ||