Please login if you have access to particular applications.
Event details
Voyage: | IN2019_T01 [details] | ||
Subject: | flow adjustment valve referbished | ||
When: | 2019-05-04 23:36:43 UTC - entered into Elog/Everlog as 2019-05-04 23:36:43 | ||
Log: | in2019_t01_Atmospherics | ||
Equipment: | CCNC
Cloud Condensation Nuclei Counter [view] | ||
Position: |
Latitude: -37.673 Longitude: 129.43067
- entered into Elog as 37 40.38 S, 129 25.84 E | ||
Action: | Start | ||
| |||
Comments: | valve removed and cleaned in the ultrasonic bath o-ring replaced, had visible sign of leaking pas seal surface-most likely cause of low flow problem flow ratio now 9.5 hepa zero test concestration now less than 1 - very good |