Please login if you have access to particular applications.
Event details
Voyage: | IN2019_V06 [details] |
Subject: | Flow check |
When: | 2019-12-18 01:42:28 UTC - entered into Elog/Everlog as 2019-12-18 01:42:28 |
Log: | in2019_v06_Atmospherics |
Equipment: | MAAP
Thermo Scientific Model 5012 MultiAngle Atmospheric Photometer [view] |
Position: | Position not recorded in Elog/Everlog. |
Action: | Maintenance |
System nominal (MAAP reported) flow: | 448Nl/h, 500l/h |
Gilibrator flow (average of 5): | 8.008 L/min |
Comments: | Tested flow with Gilibrator with high flow cell 1810008-H. Air drawn from Lab, plumbed into MAAP input System nominal (MAAP reported) flow: 448Nl/h, 500l/h Gilibrator flow (average of 5): 8.008 L/min |