Event details
Voyage: | IN2019_V06 [details] |
Subject: | MFC swapped for upstream ssensor |
When: | 2019-11-15 22:54:38 UTC - entered into Elog/Everlog as 2019-11-15 22:54:38 |
Log: | in2019_v06_Atmospherics |
Equipment: | Nephelometer
Atmospheric nephelometer [view] |
Position: |
Latitude: -11.20667 Longitude: 133.4435
- entered into Elog as 11 12.4 S, 133 26.61 E |
Action: | Change |
Parameters before change (upstream sensor): | |
- Control: | 6.00 LPM Vol Flow |
Parameters after change: | |
Comments: | MFC 179325 removed - has upstream sensor. MFC 179324 installed - has downstream sensor. 179325 will be modifies for downstream sensor and reinstalled shortly. The MFC for the bypass ( 179327) had a sensor downstream of the valve. Parameters before change (upstream sensor): - Control: 5.00 SLPM Mass Flow - Vol Flow = 6.13 LPM - Mass FLow = 5.00 SLPM - pressure = 848mBarA
Parameters after change: - Control: 6.00 LPM Vol Flow - Vol Flow = 6.00 LPM - Mass FLow = 5.54 SLPM - Pressure = 951 mBarA |