Please login if you have access to particular applications.
Event details
Voyage: | IN2019_V06 [details] | ||
Subject: | MAAP flow rate reduced to 500l/hr | ||
When: | 2019-11-12 20:13:41 UTC - entered into Elog/Everlog as 2019-11-12 20:13:41 | ||
Log: | in2019_v06_Atmospherics | ||
Equipment: | MAAP
Thermo Scientific Model 5012 MultiAngle Atmospheric Photometer [view] | ||
Position: |
Latitude: -11.203 Longitude: 133.43117
- entered into Elog as 11 12.18 S, 133 25.87 E | ||
Action: | Change | ||
| |||
Comments: | Based on testing, the MAAP flow rate has been reduced to 500L/hr, from1000L/hr. To ensure VSCC is at correct flow (1000L/hr), a bypass is running at 7.33l/min, and the sample line monitoring (t,rh,p) at 1l/min 7.33+1+8.33 = 16.66l/min = 1000L/hr Sample humidity 33.7%, Dew Point 8.75C. External air dew point = 23.5C |