Event details
| Voyage: | IN2019_V06 [details] |
| Subject: | Flow rate change tests |
| When: | 2019-11-12 19:47:31 UTC - entered into Elog/Everlog as 2019-11-12 19:47:31 |
| Log: | in2019_v06_Atmospherics |
| Equipment: | MAAP
Thermo Scientific Model 5012 MultiAngle Atmospheric Photometer [view] |
| Position: |
Latitude: -11.19983 Longitude: 133.4305
- entered into Elog as 11 11.99 S, 133 25.83 E |
| Action: | Maintenance |
| Flow rate: | |
| Comments: | Sample humidity 50% at start of test Flow rate: @1000l/hr, reg 33.7%, Stable @800l/hr, reg 20.9%, stable speed controller adjusted (orifice reduced) @800l/hr, reg 33.9%, stable @750l/hr, reg 30.6%, stable @700l/hr, reg 25.9%, stable @650l/hr, reg 18.0%, stable speed controller adjusted (orifice reduced) @650l/hr, reg 34.0%, stable sample humidity 39% @600l/hr, reg 28.7%, stable speed controller adjusted (orifice reduced) @550l/hr, reg 34.0%, stable @500l/hr, reg25.5%, stable sample humidity 33.7% Flow rate left at 500l/hr |