Please login if you have access to particular applications.
Event details
Voyage: | IN2019_V06 [details] | ||
Subject: | flow rate change testing | ||
When: | 2019-11-12 18:15:32 UTC - entered into Elog/Everlog as 2019-11-12 18:15:32 | ||
Log: | in2019_v06_Atmospherics | ||
Equipment: | MAAP
Thermo Scientific Model 5012 MultiAngle Atmospheric Photometer [view] | ||
Position: |
Latitude: -11.00933 Longitude: 133.3165
- entered into Elog as 11 00.56 S, 133 18.99 E | ||
Action: | Maintenance | ||
| |||
Comments: | @flow = 1000l/hr, reg = 33.7%. Stable @flow = 900l/hr, reg = 29.2%. Stable @flow = 800l/hr, reg = 21.8%. Stable @flow = 700l/hr, reg = ---------- Unstable @flow = 750l/hr, reg <20% Unstable @flow = 790l/hr, reg = 19.6%. Stable @flow = 780l/hr, reg = 19.2%. Stable Test completed @ 0635UTC |