Please login if you have access to particular applications.
Event details
| Voyage: | IN2019_V06 [details] | ||
| Subject: | Software hasn't been communicating with the CPC since the nozzle clean. | ||
| When: | 2019-11-09 04:53:40 UTC - entered into Elog/Everlog as 2019-11-09 04:53:40 | ||
| Log: | in2019_v06_Atmospherics | ||
| Equipment: | CPC 3772
Condensation Particle Counter (TSI 3772) [view] | ||
| Position: |
Latitude: -13.08483 Longitude: 125.28883
- entered into Elog as 13 05.09 S, 125 17.33 E | ||
| Action: | Outage | ||
| |||
| Comments: | Ian is logged on remotely to get it going again. When I Plugged the serial plug back into the MFC, it sparked a bit. Ian has fixed it remotely by sending the instrument a "SSTART,0" and "SSTART,1" set of commands to wake it back up again. | ||