Voyage: | IN2019_V06 [details] | ||
Subject: | Took MAAP off main inlet (closed manifold tap) to test flow rate as source of error | ||
When: | 2019-10-23 03:41:51 UTC - entered into Elog/Everlog as 2019-10-23 03:41:51 | ||
Log: | in2019_v06_Atmospherics | ||
Equipment: | MAAP
Thermo Scientific Model 5012 MultiAngle Atmospheric Photometer [view] | ||
Position: |
Latitude: -14.29033 Longitude: 121.236
- entered into Elog as 14 17.42 S, 121 14.16 E | ||
Action: | Fault | ||
| |||
Comments: | Seems turning the flowrates back up to 1000 m3/hr removed most of the errors. Not sure yet about the filter change error. I& 39;ve reconnected back to the inlet, running at 16.67 LPM and restarted the machine. This seems to have cleared all the errors for now. |