Please login if you have access to particular applications.
Event details
Voyage: | IN2019_V06 [details] | ||
Subject: | Orifice blocked (fixed) | ||
When: | 2019-11-14 00:47:33 UTC - entered into Elog/Everlog as 2019-11-14 00:47:33 | ||
Log: | AIRBOX_in2019_v06 | ||
Equipment: | SMPS
nano Scanning Mobility Particle Sizer (Grimm) [view] | ||
Position: |
Latitude: -11.20683 Longitude: 133.43067
- entered into Elog as 11 12.41 S, 133 25.84 E | ||
Action: | Maintenance | ||
| |||
Comments: | Poly flow was 0.21, adjusted sheath flow knob until 0.31. This lowered sheath flow to 1.66 (2.42 on classifier). After attempting to correct this (with no success) we found the critical orifice was blocked. We cleaned it and found sheath flow had risen to 2.9. Adjusted excess and sheath flows until all were normal again |