Voyage: IN2019_V06       [details]
Subject: Drying out CPC
When: 2019-11-05 10:02:54 UTC

- entered into Elog/Everlog as 2019-11-05 10:02:54
Log: AIRBOX_in2019_v06
Equipment: CPC3772

Condensation Particle Counter (TSI 3772) [view]
Position: Latitude: -13.79867 Longitude: 121.95817

- entered into Elog as 13 47.92 S, 121 57.49 E
Action: Maintenance
Comments:

I have put a dryer on the inlet of the CPC and taken it of line. CPC counts 387 cm-3; SMPScount before the CPC was disconected is 1.14e3cm-3

After the CPCrunning through dryer it counts 328cm-3