Please login if you have access to particular applications.
Event details
Voyage: | IN2020_V01 [details] |
Subject: | Span check ran longer than usual (Nephelometer Interface firmware bug) |
When: | 2020-03-02 01:00:00 UTC - entered into Elog/Everlog as 2020-03-02 01:00:00 |
Log: | in2020_v01_Atmospherics |
Equipment: | Nephelometer
Atmospheric nephelometer [view] |
Position: |
Latitude: -33.72583 Longitude: 101.87617
- entered into Elog as 33 43.55 S, 101 52.57 E |
Action: | Fault |
After the scheduled span check a firmware bug in the Nepheloemter Interface caused the Nephelometer to run another span check for a longer than usual time. ;Exhaust was detected during this period, so the Nepheloemer ;should have been set to zero check instead. The bug was fixed at ;2020-03-02 01: | 05.280 UTC. |
Comments: | After the scheduled span check a firmware bug in the Nepheloemter Interface caused the Nephelometer to run another span check for a longer than usual time. Exhaust was detected during this period, so the Nepheloemer should have been set to zero check instead. The bug was fixed at 2020-03-02 01:32:05.280 UTC. |