Survey | When (UTC) | Latitude | Longitude | Instrument | Identifier | Action | Log | Subject / Comments |
---|---|---|---|---|---|---|---|---|
IN2017_T01 | 2017-09-29 17:47 [view ] | -13.1332 | 147.8770 | Nephelometer [view] | Stop | in2017_t01 | Nephelometer stopped due to high RH | |
IN2017_V04 | 2017-09-06 11:27 [view ] | -33.0809 | 154.5588 | Nephelometer [view] | maintenance | in2017_v04 | Line interruption Line interrupted for T&RH sensor install. May also affect CPC and SMPS | |
IN2018_T01 | 2018-04-12 23:30 [view ] | -31.3683 | 154.1238 | Nephelometer [view] | Stop | in2018_t01 | Stopped due to high humidity | |
IN2018_T02 | 2018-05-20 19:47 [view ] | -43.0200 | 147.3830 | Nephelometer [view] | Stop | Atmospherics_in2018_t02 | End of voyage shutdown | |
IN2018_V01 | 2018-02-03 00:42 [view ] | -64.9980 | 145.4994 | Neph [view] | Atmospherics in2018_v01 | Running full cal Measurements seem to be below zero... | ||
IN2018_V01 | 2018-02-03 03:55 [view ] | -65.0902 | 146.2329 | Neph [view] | Atmospherics in2018_v01 | Running zero check | ||
IN2018_V06 | 2018-11-27 00:25 [view ] | -44.2872 | 147.1577 | Nephelometer [view] | Maintenance | in2018_v06_Atmospherics | filter and cal routine Looked at the cal data from the voyage thus far. Seems like the zero checks are a bit less stable than they used to be so I& 39;ve put the old filter back in. Also noticed there were no span checks taking place. I& 39;ve changed the cal routine to do zero and span checks every 24hrs. | |
IN2018_V06 | 2018-12-12 01:41 [view ] | -44.2392 | 147.4655 | Nephelometer [view] | Maintenance | in2018_v06_Atmospherics | MFC connection testing Did some tests to confirm best vacuum source for system. Chose the weak vacuum - no advantage to strong vacuum | |
IN2018_V06 | 2018-12-13 21:17 [view ] | -44.3530 | 147.2775 | Nephelometer [view] | Maintenance | in2018_v06_Atmospherics | Testing serial control Sent multidrop command **0J4 to start a zero check. Testing to see if we can control the device to avoid ship exhaust contamination | |
IN2018_V06 | 2018-12-13 21:46 [view ] | -44.3522 | 147.2767 | Nephelometer [view] | Maintenance | in2018_v06_Atmospherics | Testing serial control - complete Completed the test. Returned the device to normal state with the command **0J0 This test confirmed that: - multidrop port commands do not interfere with data output on the service port - I can set the device state to zero check or normal monitoring. presumably can do other states similarly - device will stay in the assigned state until further instruction | |
IN2018_V06 | 2018-12-15 02:38 [view ] | -43.6125 | 147.4400 | Nephelometer [view] | Observation | in2018_v06_Atmospherics | testing status monitoring currently testing a status monitoring script, for use with a future exhaust shutdown/protection. Communicating via Multidrop port. Shouldn& 39;t affect neph operation, but any anomalies may be attributable to this | |
IN2018_V06 | 2018-12-15 05:11 [view ] | -43.1383 | 147.4657 | Nephelometer [view] | Maintenance | in2018_v06_Atmospherics | testing exhaust bypass Tested the exhaust bypass algorithm between 0500 and 0515. This set the neph to zero check mode when the exhaust algorithm detected contamination. Test showed this can work. still need to work out how to integrate daily span and zero checks. | |
IN2018_V06 | 2018-12-15 17:51 [view ] | -42.1940 | 148.4982 | Nephelometer [view] | Change | in2018_v06_Atmospherics | Neph Control script started A script to control the nephelometer state has been started. This should: set the neph to zero check if exhaust is present set zero check for 30 mins from 0000-0030UTC set span check for 30 mins from 0030-0100UTC lets see how it goes... | |
IN2018_V06 | 2018-12-15 21:09 [view ] | -41.7477 | 148.6658 | Nephelometer [view] | Change | in2018_v06_Atmospherics | time set time was incorrect, so was set to UTC | |
IN2018_V06 | 2018-12-15 21:10 [view ] | -41.7455 | 148.6682 | Nephelometer [view] | Change | in2018_v06_Atmospherics | regular cal checks disabled I have disabled the instrument& 39;s scheduled zero and span checks, as this is now managed by the exhaust bypass control software | |
IN2018_V06 | 2018-12-17 00:52 [view ] | -41.2332 | 148.7495 | Nephelometer [view] | Fault | in2018_v06_Atmospherics | Bug in exhaust monitor code Found a bug in the exhaust monitor code which was preventing the scheduled span and zero checks from operating properly. Hopefully have fixed this. | |
IN2018_V06 | 2018-12-17 19:23 [view ] | -41.5648 | 148.7778 | Nephelometer [view] | Maintenance | in2018_v06_Atmospherics | restarted control script | |
IN2018_V08 | 2019-01-01 05:02 [view ] | -36.2840 | 158.8133 | Nephelometer [view] | Start | in2018_v08_Atmospherics | Restarted The Nephelometer was displaying "Instrument in Warmup" on its display. I power cycled it to see if the message goes away. I also stopped the "Neph Exhaust" script that Ian had set up. | |
IN2019_E01 | 2019-07-23 01:50 [view ] | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Nephelometer 12 month maintenance including: - Exhaust ball valve cleaning - inlet ball valve was clean, but exhaust ball valve was quite dirty. - Optic cell clean performed - this was quite clean. - Full optical bench clean. A small amount of vac grease was applied to o-rings in optic chamber tubes. - Inspected the exhaust flow DFU - all OK, none replaced. - Batteries were replaced. - Pneumatics were cleaned After reassembly, a leak check was performed - Failed Isolated the external plumbing from the leak test - still failed. Discovered tubing to the pressure sensor was pinched during optic bench reinstallation - test still failed. Found a tube completely disconnected on the output of the zero pump, which means the zero tests for the foreseeable past are crap, but will need to look into the past zeros to verify data quality. This is an inherent problem with the design of these connections. | |||
IN2019_E01 | 2019-07-28 05:52 [view ] | -10.1250 | 135.9800 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Running overnight zero, starting at 05:50am UTC | |
IN2019_E01 | 2019-07-29 10:20 [view ] | -10.7585 | 141.5920 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Zero check stopped, started back on ambient | |
IN2019_E01 | 2019-07-30 01:02 [view ] | -9.2247 | 143.8993 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Added missing insulation to inlet pipe that was missed during servicing. | |
IN2019_E01 | 2019-07-30 04:21 [view ] | -9.5032 | 144.4978 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Temperature, BaroPress, RH sensor calibration Audited sample temperature, Barometric pressure, and sample RH sensors against external probes (about 2hrs ago). All passsed, no adjustment performed. Parameter Nephelometer Calibrated reference Temperature 25.69 26.34 Rel Humidity 44.775 44.93 Pressure 1010.1 1009 | |
IN2019_E01 | 2019-07-30 04:24 [view ] | -9.5110 | 144.5042 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | leak check performed instrument leak check, from software settings (exhaust plumbed to inlet). all passed. | |
IN2019_E01 | 2019-07-30 04:30 [view ] | -9.5265 | 144.5167 | Nephelometer [view] | Start | in2019_e01_Atmospherics | Zero check | |
IN2019_E01 | 2019-07-30 05:12 [view ] | -9.6367 | 144.6057 | Nephelometer [view] | Start | in2019_e01_Atmospherics | Span check | |
IN2019_E01 | 2019-07-30 06:18 [view ] | -9.7598 | 144.7603 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Full calibration activated | |
IN2019_E01 | 2019-07-30 07:34 [view ] | -9.7907 | 144.7482 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Span check post-adjust Span check | |
IN2019_E01 | 2019-07-30 08:11 [view ] | -9.7943 | 144.7415 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Zero check Post-adjust Zero Check | |
IN2019_E01 | 2019-07-30 08:44 [view ] | -9.7860 | 144.7487 | Nephelometer [view] | Observation | in2019_e01_Atmospherics | Zero check Zero cal/check does not appear to be working proplery. Cal ball valves did not close, cal control box did not work, zero air pump did not switch on. Will investigate tomorrow. | |
IN2019_E01 | 2019-07-30 09:24 [view ] | -9.8572 | 144.7975 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | Zero Check test tested zero check functionality via remote terminal /udpterminal | |
IN2019_E01 | 2019-07-31 03:24 [view ] | -11.9405 | 145.4648 | Nephelometer [view] | Maintenance | in2019_e01_Atmospherics | flow audit Neph disconnected from sample line 02:30 - 03:20hrs for flow calibration. Flows measured at inlet of Neph. Gilib (initial) = 5.121LPM (n=5). Alicat set = 5.000slpm Gilib (final) = 5.019LPM (n=5). Alicat set= 4.950slpm, read= 4.950slpm/14.76(v)lpm. | |
IN2019_T01 | 2019-03-06 17:41 [view ] | -42.8863 | 147.3388 | Nephelometer [view] | Observation | in2019_t01_Atmospherics | CO2 Cal Tank Empty CO2 tank was noted as empty, unsure when exactly this occured. High side of the regulator is showing 0.0 Low side is showing <40psi | |
IN2019_V03 | 2019-05-14 12:30 [view ] | -32.4402 | 115.0838 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | Span check test /span check activated to test new co2 flow meter | |
IN2019_V03 | 2019-05-17 09:52 [view ] | -39.4982 | 110.0043 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | Full calibration activated | |
IN2019_V03 | 2019-05-17 11:22 [view ] | -39.4605 | 109.9713 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | zero check following cal | |
IN2019_V03 | 2019-05-17 12:06 [view ] | -39.4528 | 109.9675 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | Normal operation started | |
IN2019_V03 | 2019-05-24 08:40 [view ] | -29.0007 | 110.0015 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | Zero adjust started | |
IN2019_V03 | 2019-05-24 08:46 [view ] | -29.0018 | 110.0017 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | vacuum outage Brief vacuum outage due to maintenance on vacuum line. | |
IN2019_V03 | 2019-05-24 08:49 [view ] | -29.0020 | 110.0017 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | Zero check started | |
IN2019_V03 | 2019-05-24 08:55 [view ] | -29.0033 | 110.0018 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | Restarted | |
IN2019_V03 | 2019-05-24 08:55 [view ] | -29.0033 | 110.0018 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | vacuum stopped temporarly | |
IN2019_V03 | 2019-05-25 10:05 [view ] | -27.4583 | 109.9738 | Nephelometer [view] | Maintenance | in2019_v03_Atmospherics | Restarted due to work on vacuum | |
IN2019_V03 | 2019-06-12 08:17 [view ] | -28.4530 | 113.4793 | Nephelometer [view] | Stop | in2019_v03_Atmospherics | Shut down for dry dock | |
IN2019_V06 | 2019-10-22 01:52 [view ] | -13.9562 | 121.3962 | Nephelometer [view] | in2019_v06_Atmospherics | Exhaust detection min wind threshold changed from 10 to 3 kt This will affect the exhaust zeroing. | ||
IN2019_V06 | 2019-11-07 06:00 [view ] | -13.7993 | 121.9553 | Nephelometer [view] | Observation | in2019_v06_Atmospherics | Neph hasn't done its calibration the past few days - CO2 tank has run out. Will is going to replace it this shift. | |
IN2019_V06 | 2019-11-07 06:41 [view ] | -13.7987 | 121.9577 | Nephelometer [view] | Maintenance | in2019_v06_Atmospherics | co2 tank replaced - co2 tank depleted, spare connected, and leak checked - accumulative flow meter reading 456 - accumulative meter reset | |
IN2019_V06 | 2019-11-15 22:54 [view ] | -11.2067 | 133.4435 | Nephelometer [view] | Change | in2019_v06_Atmospherics | MFC swapped for upstream ssensor MFC 179325 removed - has upstream sensor. MFC 179324 installed - has downstream sensor. 179325 will be modifies for downstream sensor and reinstalled shortly. The MFC for the bypass ( 179327) had a sensor downstream of the valve. Parameters before change (upstream sensor): - Control: 5.00 SLPM Mass Flow - Vol Flow = 6.13 LPM - Mass FLow = 5.00 SLPM - pressure = 848mBarA Parameters after change: - Control: 6.00 LPM Vol Flow - Vol Flow = 6.00 LPM - Mass FLow = 5.54 SLPM - Pressure = 951 mBarA | |
IN2019_V06 | 2019-11-16 00:11 [view ] | -11.2057 | 133.4437 | Nephelometer [view] | Maintenance | in2019_v06_Atmospherics | MFC reinstalled MFC 179325 reinstalled. Parameters after change: - Control: 6.00 LPM Vol Flow - Vol Flow = 6.00 LPM - Mass FLow = 5.64 SLPM - Pressure = 968 mBarA | |
IN2019_V06 | 2019-11-18 17:41 [view ] | -11.8435 | 131.5385 | Nephelometer [view] | Change | in2019_v06_Atmospherics | Flow rate reduced to 5LPM (Vol) Flow rate reduced from 6LPM to 5 LPM. Sample line humidity was starting to exceed 40% at times | |
IN2019_V07 | 2019-04-18 01:18 [view ] | -38.5988 | 147.5725 | Nephelometer [view] | Maintenance | Atmospherics_in2019_v07 | Zero Adjust | |
IN2019_V07 | 2019-04-18 01:18 [view ] | -38.5988 | 147.5725 | Nephelometer [view] | Maintenance | in2019_v07_Atmospherics | Zero Adjust | |
IN2020_E01 | 2020-08-05 19:54 [view ] | -42.9715 | 147.3807 | Nephelometer [view] | Outage | in2020_e01_Atmospherics | ||
IN2020_V01 | 2020-01-30 20:38 [view ] | -54.4142 | 81.9503 | Nephelometer [view] | Maintenance | in2020_v01_Atmospherics | Stopped to clean optical chamber Nothing came out with triton X solution, a lot came out with IPA wipes | |
IN2020_V01 | 2020-01-31 02:28 [view ] | -54.6427 | 82.5350 | Nephelometer [view] | Maintenance | in2020_v01_Atmospherics | Running post clean leak check Leak check didn& 39;t complete first few attempts (just hanging for 15+ minutes) tried power cycling neph and calibration controller, still didn& 39;t complete. Stopped NephControl script and ran leak test, it passed | |
IN2020_V01 | 2020-01-31 02:42 [view ] | -54.6560 | 82.5397 | Nephelometer [view] | Maintenance | in2020_v01_Atmospherics | Zero noise check start zero noise started 2h duration | |
IN2020_V01 | 2020-01-31 04:49 [view ] | -54.5815 | 82.6288 | Nephelometer [view] | Maintenance | in2020_v01_Atmospherics | Full calibration started full calibration started min 25min max 30min | |
IN2020_V01 | 2020-01-31 05:33 [view ] | -54.6513 | 82.7347 | Nephelometer [view] | Start | in2020_v01_Atmospherics | Start NephControl script after calibration | |
IN2020_V01 | 2020-01-31 05:47 [view ] | -54.6767 | 82.7738 | Nephelometer [view] | Maintenance | in2020_v01_Atmospherics | Started calibration Calibration didn& 39;t finish properly, started it again | |
IN2020_V01 | 2020-02-26 01:34 [view ] | -32.9015 | 97.6633 | Nephelometer [view] | Change | in2020_v01_Atmospherics | Moving NephControl scrip functionality to "Nephelometer Interface" Stopped NephControl script on Aerosol GP PC. Connected Multidrop and Service Serial ports to new Nephelometer Interface box (0194 board). Nephelometer data, control commands, acknowledgements of control commands, and CO2 flow rates are being sent to techsas on port 51503. IP address of Nepheloemeter Interface 0194 board is 150.229.233.244 Changes were made while exhaust was being detected, so as not to lose any good data | |
IN2020_V01 | 2020-02-27 00:00 [view ] | -32.0442 | 98.6372 | Nephelometer [view] | Fault | in2020_v01_Atmospherics | Zero checks didn't run properly (Nephelometer Interface firmware bug) On the 2020-02-27 and 2020-02-28 the zero checks didn& 39;t run properly due to a firmware bug. The bug was fixed on the 2020-02-28. | |
IN2020_V01 | 2020-02-28 00:00 [view ] | -32.8720 | 99.9995 | Nephelometer [view] | Fault | in2020_v01_Atmospherics | Zero checks didn't run properly (Nephelometer Interface firmware bug) On the 2020-02-27 and 2020-02-28 the zero checks didn& 39;t run properly due to a firmware bug. The bug was fixed on the 2020-02-28. | |
IN2020_V01 | 2020-03-02 01:00 [view ] | -33.7258 | 101.8762 | Nephelometer [view] | Fault | in2020_v01_Atmospherics | Span check ran longer than usual (Nephelometer Interface firmware bug) After the scheduled span check a firmware bug in the Nepheloemter Interface caused the Nephelometer to run another span check for a longer than usual time. Exhaust was detected during this period, so the Nepheloemer should have been set to zero check instead. The bug was fixed at 2020-03-02 01:32:05.280 UTC. | |
IN2020_V01 | 2020-03-03 03:33 [view ] | -32.8690 | 102.7092 | Nephelometer [view] | Change | in2020_v01_Atmospherics | Set Nephelometer to correct time Nephelometer real-time clock has displaying the incorrect time since dry dock, I updated it to reflect the current UTC time. I also updated the Nepheloemeter interface to set the time when starting up. The time immediately before change: 10/08/2003 21:33:45 The time immediatelly after change: 03/03/2020 03:33:20 | |
IN2020_V06 | 2020-10-15 17:25 [view ] | -54.6003 | 158.9608 | Nephelometer [view] | Start | in2020_v06_Atmospherics | Restarted Sensor Interface I restarted the sensor interface because the nepholometer has been saying "Warming Up" since we left. Seeing if that message will disappear. | |
IN2020_V06 | 2020-10-15 23:47 [view ] | -54.3365 | 159.1801 | Nephelometer [view] | Outage | in2020_v06_Atmospherics | 0194 stopped to test system warmup The neph keeps reporting "instrument in warmup". I& 39;m testing ahypothesis that the 0194 is preventing the neph from doing it& 39;s normal warmup procedure I have shut down the 0194 application code for 15 minutes, and will then restart it. | |
IN2020_V06 | 2020-10-20 02:33 [view ] | -55.1128 | 159.1562 | Nephelometer [view] | Outage | in2020_v06_Atmospherics | Turned off for 10minutes to fix warmup message on display. | |
IN2021_T01 | 2021-06-16 23:42 [view ] | -11.8255 | 145.0505 | Nephelometer [view] | Fault | Atmospherics | air leak into the smps sample 16/6 23:26 - I just found that the air inlet tube was not connected properly. air has been sucked into the sampling system since I connected up the SMPS on 13/6 18:55. This will have affected the SMPS, and CPCs. It is possible that it will have affected the.nephelometer and CCNC. It is possible, but I think unlikely, that it also affected the ozone and greenhouse gas system, depending on the interactions in the manifold and sample lines. | |
IN2021_T01 | 2021-06-21 05:25 [view ] | -10.7505 | 134.5233 | Nephelometer [view] | Maintenance | Atmospherics | Swap out CO2 cylinder and regulator (cylinder fitting swap) | |
IN2021_V01 | 2021-02-08 21:34 [view ] | -62.4927 | 88.1003 | Nephelometer [view] | Outage | Atmospherics | Nephelometer disconnected from inlet (inlet plugged) Nephelometer disconnected from the inlet to perform a light source check. The inlet was plugged to not contaminate other instruments. | |
IN2021_V01 | 2021-02-08 22:47 [view ] | -62.4585 | 87.6815 | Nephelometer [view] | Change | Atmospherics | Reconnected to inlet after light source check and optical clean | |
IN2021_V01 | 2021-02-08 22:55 [view ] | -62.4588 | 87.6235 | Nephelometer [view] | Maintenance | Atmospherics | Full calibration started after optical chamber clean | |
IN2021_V01 | 2021-02-09 02:41 [view ] | -62.4745 | 86.0873 | Nephelometer [view] | Maintenance | Atmospherics | Restarting full cal, used wrong settings the first time | |
IN2021_V01 | 2021-03-12 04:52 [view ] | -65.0522 | 80.1270 | Nephelometer [view] | Stop | Atmospherics | Stopping Nephelometer for optical chamber clean | |
IN2021_V01 | 2021-03-12 05:21 [view ] | -65.0248 | 80.2550 | Nephelometer [view] | Maintenance | Atmospherics | Disconnected inlet and plugged (performing leak check) | |
IN2021_V01 | 2021-03-12 05:34 [view ] | -65.0187 | 80.3133 | Nephelometer [view] | Maintenance | Atmospherics | Leak test successful, inlet reconnected | |
IN2021_V01 | 2021-03-12 06:14 [view ] | -64.9802 | 80.4933 | Nephelometer [view] | Maintenance | Atmospherics | Zero Adjust complete | |
IN2021_V01 | 2021-03-12 06:14 [view ] | -64.9798 | 80.4947 | Nephelometer [view] | Start | Atmospherics | ||
IN2021_V04 | 2021-06-29 05:15 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Maintenance | Atmospherics | Pre-maintainance leak check succesful | |
IN2021_V04 | 2021-06-29 05:47 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Maintenance | Atmospherics | Measurement cell cleaned | |
IN2021_V04 | 2021-06-29 05:56 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Maintenance | Atmospherics | Leak check successful after measurement cell clean | |
IN2021_V04 | 2021-06-29 05:58 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Maintenance | Atmospherics | Zero noise test - connected HEPA filter to inlet | |
IN2021_V04 | 2021-06-29 07:15 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Change | Atmospherics | Exhaust checks disabled temporarily Exhaust checks disabled for zero noise test and calibration | |
IN2021_V04 | 2021-06-29 11:44 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Change | Atmospherics | Inlet reconnected to manifold after zero test | |
IN2021_V04 | 2021-06-29 11:48 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Maintenance | Atmospherics | Full calibration started | |
IN2021_V04 | 2021-06-29 16:31 [view ] | -12.4718 | 130.8468 | Nephelometer [view] | Change | Atmospherics | Restarting exhaust checks after full calibration | |
IN2021_V04 | 2021-07-16 06:16 [view ] | -11.3733 | 104.4913 | Nephelometer [view] | Change | Atmospherics | Changed the Nephelometer Interface firmware to v1.03 This firmware update changes the way exhaust detection works - Neph will now enter zero check as soon as exhaust is detected and will only change back to normal when exhaust has not been detected continuously for 120 seconds. | |
IN2021_V04 | 2021-07-19 12:10 [view ] | -20.4683 | 107.9732 | Nephelometer [view] | Change | Atmospherics | Exhaust checks disabled for leak testing | |
IN2021_V04 | 2021-07-19 13:18 [view ] | -20.6615 | 108.0735 | Nephelometer [view] | Change | Atmospherics | Exhaust checks turned back on after leak test | |
IN2021_V04 | 2021-07-29 00:08 [view ] | -42.9888 | 147.3773 | Nephelometer [view] | Stop | Atmospherics | Shutdown Neph turned off. Interface turned off. Gas bottle closed. |