Survey When (UTC) LatitudeLongitude InstrumentIdentifierActionLogSubject / Comments
IN2015_C01 2015-10-31 01:02 [view ] -36.1344 132.7825 Aerosol Lab [view] restarted in2015_c01 Cycled power on equipment to locate earth fault for ship electrician. Cycled power on Radon detector, O3 monitors, air inlet controller, air inlet monitor, nephelometer and exhaust pump.
IN2016_V02 2016-04-09 22:00 [view ] -51.9024 148.5204 Aerosol Lab [view] Repair in2016_v02 Removed, repaired, and replaced peristaltic pump for manifold ~22:00 -23:00
IN2016_V03 2016-04-27 01:25 [view ] -45.2368 149.2368 Aerosol Lab [view] Start in2016_v03 Aerosol Manifold sensor Interface Aerosol Manifold sensor interface was found switched off, so I switched it on. Probably switched off during port period.
IN2016_V03 2016-04-29 22:28 [view ] -55.5099 158.2691 Aerosol Lab [view] Air Sampling in2016_v03 peristaltic pump swapped The water trap peristaltic pump has been swapped out. Some lab air may have contaminated the sample air for instruments connected to the main air sample line in the last 15 minutes.
IN2016_V03 2016-05-27 16:32 [view ] -41.9411 179.0248 Aerosol Lab [view] Stop in2016_v03 Manifold temp/humidity sensor interface turned off and removed for modifications
IN2016_V03 2016-05-27 21:00 [view ] -42.0291 -179.7247 Aerosol Lab [view] Start in2016_v03 Manifold temp/humidity sensor interface modified and turned on
IN2016_V05 2016-09-30 11:33 [view ] -20.9935 150.1979 Aerosol Lab [view] Air Sampling in2016_v05 Vacuum interrupted 11:33 30/09/16 UTC - 11:35 30/09/16 UTC Installed Nanometer Aerosol Sampler on starboard bench in the aerosol lab. Interrupted the vacuum line
IN2016_V05 2016-10-01 21:31 [view ] -22.4912 152.8668 Aerosol Lab [view] CPC in2016_v05 CPC restarted CPC restarted. Serial splitter installed
IN2016_V05 2016-10-02 01:13 [view ] -22.0640 153.0356 Aerosol Lab [view] Air Sampling in2016_v05 disruption to aerosol manifold The NOx tube was disconnected from the manifold and capped. Lab air will have entered the system for ~30s
IN2016_V05 2016-10-02 22:25 [view ] -20.8275 153.0663 Aerosol Lab [view] Air Sampling in2016_v05 Line opened - test filter flow 22:23 02/10/2016 UTC - 22:25 02/10/2016 UTC. testing flows on QUT filter setup.
IN2016_V05 2016-10-03 05:50 [view ] -20.8207 153.0922 Aerosol Lab [view] Air Sampling in2016_v05 Attached PSM to manifold 05:50 3/10/2016 - 5:52 03/10/2016 UTC Opened manifold to attach PSM, installation was quick.
IN2016_V05 2016-10-03 10:05 [view ] -20.8160 153.1098 Aerosol Lab [view] Air Sampling in2016_v05 Flow checks on QUT filter line 10:05 03/10/2016 UTC - 10:10 03/10/2016 UTC. Checked flows for QUT filter, this should on;ly influence instruments downstream -> TSI SMPS, CPC3776, TEM and possibly Neph.
IN2016_V05 2016-10-04 04:37 [view ] -20.3707 152.2956 Aerosol Lab [view] Software Update in2016_v05 Pump switch controller code updated Pump switch controller code updated to correct bug which was not switching the outlets properly based on wind bearing.
IN2016_V05 2016-10-07 05:44 [view ] -18.5795 146.9728 Aerosol Lab [view] Start in2016_v05 flowmeter logger started Logging code for the TSI4100 flowmeter has been installed on the starboard bench PC and is logging data at 1Hz. data fields are flow (standard units), temperature and pressure.
IN2016_V05 2016-10-12 00:30 [view ] -18.4443 146.6754 Aerosol Lab [view] Air Sampling in2016_v05 Removed inlet drier 00:30 12/10/2016 UTC - 00:45 12/10/2016 UTC. Trying to resolve variable CPC/SMPS concentrations when the filter pump turns on/off. Removed inlet drier, suspect it could be leaking (turned out not to be the case -> there was an issue with back flow through the Neph).
IN2016_V05 2016-10-12 02:10 [view ] -18.4421 146.6804 Aerosol Lab [view] Air Sampling in2016_v05 Installed RH probe 02:10 12/10/2016 UTC - 02:15 12/10/2016 UTC. Installed RH probe on SMPS line. Needed to disrupt vacuum line, and open inlet at TSI SMPS.
IN2016_V05 2016-10-12 07:21 [view ] -18.5164 146.7892 Aerosol Lab [view] Air Sampling in2016_v05 Returned inlet drier 7:17 12/10/2016 UTC - 7:21 12/10/2016 UTC. Returned the inlet drier to the line - disconnected the inlet along the aerosol instrumentation line. I have also change the filter between the silica canisters and the nation drier. The old filter was a bit blocked.
IN2016_V05 2016-10-12 08:05 [view ] -18.5674 146.8794 Aerosol Lab [view] Air Sampling in2016_v05 Changing over vacuum lines 8:05 12/10/2016 UTC - 8:15 12/10/2016 UTC. Added small knf pump for vacuum to pass sample across RH probe and for inlet drier sheath. This makes the vacuum CPCs happy with their vacuum pressure.
IN2016_V05 2016-10-12 22:45 [view ] -18.6276 146.9804 Aerosol Lab [view] Air Sampling in2016_v05 Inlet opened at Neph 22:45 12/10/2016 UTC - 23:20 12/10/2016 UTC . Testing Neph inlet flows -> there was back flow through the Neph, therefore we removed the Neph from the line.
IN2016_V05 2016-10-13 01:26 [view ] -18.6246 146.9800 Aerosol Lab [view] Air Sampling in2016_v05 Inlet opened - New Neph sampling line. 01:26 13/10/2016 UTC - 1:42 13/10/2016 UTC. Added a new sampling line for the Neph directly back to the manifold. i.e. not shared with the other aerosol instrumentation and not going through the drier.
IN2016_V05 2016-10-13 01:41 [view ] -18.6246 146.9799 Aerosol Lab [view] outage in2016_v05 Interruption to air sampling line The air sampling line was interrupted during the previous hour.The nephelometer was disconnected from the starboard bench line and set up on it's own sample line from the manifold, to correct pressure fluctuations observed on the CPC's/SMPS's. Sample line is stainless steel, with conductive tubing at the ends.
IN2016_V05 2016-10-15 19:21 [view ] -18.7685 146.9174 Aerosol Lab [view] outage in2016_v05 CPC3772 offline for last hour or more The CPC 3772 has been off for maintenance for the past hour or more.
IN2016_V05 2016-10-20 02:29 [view ] -21.0243 150.1144 Aerosol Lab [view] Air Sampling in2016_v05 Check inlet flow PSM- inlet opened 02:29 20/10/2016 UTC - 02:36 20/10/2016 UTC. Disconnected inlet at manifold to check the PSM flow rate.
IN2016_V05 2016-10-21 01:00 [view ] -21.0194 150.1136 Aerosol Lab [view] Air Sampling in2016_v05 Measuring inlet flow of Neph 01:00 21/10/2016 UTC - 01: 4 0 21/10/2016 UTC - Opened sampling line at Neph to measure the inlet flow rate -> less than 1 LPM which is quite low - something wrong with the Neph pump. Note inlet wasn't open for the entire period.
IN2016_V05 2016-10-21 19:03 [view ] -21.0155 150.1101 Aerosol Lab [view] maintenance in2016_v05 SMPS butanol top up
IN2017_T01 2017-09-30 21:55 [view ] -9.4175 144.4367 Aerosol Lab [view] Change in2017_t01 T&RH sensor moved from nafion outlet to Bowie experiment outlet.
IN2017_T01 2017-10-05 05:00 [view ] -11.5548 128.3879 Aerosol Lab [view] maintenance in2017_t01 Manifold T&RH sensor modification I made a change to the manifold code to handle RH >= 100%, and tried to restart it but it has hung :( Hopefully it will kick on, but otherwise a manual restart by someone on the ship may be required.
IN2017_T02 2017-11-17 04:08 [view ] -34.7123 119.7097 Aerosol Lab [view] maintenance in2017_t02 attempted to set again ccnc flows, smps set to 1093.95 dmax, and now set to sample air
IN2017_T02 2017-11-17 20:04 [view ] -34.5373 120.3103 Aerosol Lab [view] maintenance in2017_t02 ccnc turned off and replaced filter, restarted connected to dry lab air. SMPS dmax changed to 800
IN2017_T02 2017-11-18 00:15 [view ] -34.5778 120.2235 Aerosol Lab [view] maintenance in2017_t02 CCNC, shutdown and cleared water from nafion. restarted with dried lab air.
IN2017_T02 2017-11-18 18:22 [view ] -34.5062 120.8523 Aerosol Lab [view] maintenance in2017_t02 ccnc connected to manifold air
IN2017_T02 2017-11-20 00:02 [view ] -35.0455 121.4243 Aerosol Lab [view] maintenance in2017_t02 CPC turned off and drain emptied (butanol), and radon set exhast flow tap to full open
IN2017_T02 2017-11-20 22:18 [view ] -37.7790 125.2358 Aerosol Lab [view] maintenance in2017_t02 cpc, ccnc, smps, neph line disconnected from manifold so that runs on dry lab air during high outside humidity
IN2017_T02 2017-11-21 01:55 [view ] -38.2593 125.9187 Aerosol Lab [view] maintenance in2017_t02 ccnc, smps, cpc, neph connected back to manifold air. T & RH line sensor placed closer to line.
IN2017_T02 2017-11-24 01:44 [view ] -42.1160 141.2592 Aerosol Lab [view] maintenance in2017_t02 ccnc adjusted sample flow
IN2017_T02 2017-11-24 22:08 [view ] -42.8035 145.1623 Aerosol Lab [view] maintenance in2017_t02 restart smps with dmax 1000 and 500 failed (error 256), changed to run on dry clean lab air (just smps)
IN2017_T02 2017-11-25 00:27 [view ] -43.0658 145.3715 Aerosol Lab [view] maintenance in2017_t02 smps restarted again, no 256 error, then put back on manifold air
IN2017_V03 2017-05-18 00:21 [view ] -41.7827 149.2037 Aerosol Lab [view] maintenance in2017_v03 CCNC MilliQ replaced CCNC Milliq was replaced and tested for leaks at about 2300 UTC
IN2017_V03 2017-05-27 23:48 [view ] -35.1327 151.4815 Aerosol Lab [view] maintenance in2017_v03 CCNC MilliQ replaced CCNC Milli-q replaced. zero filter tested.
IN2017_V04 2017-08-31 07:10 [view ] -26.9710 153.2280 Aerosol Lab [view] Start in2017_v04 Aerosol Instruments Running SMPS not on board this voyage
IN2017_V04 2017-09-03 01:07 [view ] -27.0040 156.0252 Aerosol Lab [view] Repair in2017_v04 Ozone monitor - change to all PFA water trap (time with potential back flow to manifold) ie between 2017-09-03 01:07:00 and 2017-09-03 01:24:00
IN2017_V04 2017-09-04 19:49 [view ] -27.8965 154.2725 Aerosol Lab [view] Software Update in2017_v04 modify aerosol lab manifold sensor code to include pump timer and water sensor (app v1.02)
IN2017_V04 2017-09-15 04:38 [view ] -35.2771 151.7722 Aerosol Lab [view] maintenance in2017_v04 Water trap sensor installed Water trap sensor installed. Manifold port open from 0345 - 0426 UTC
IN2018_T01 2018-04-12 23:30 [view ] -31.3683 154.1238 Aerosol Lab [view] Temperature Adjusted in2018_t01 Temp set to 25
IN2018_V01 2018-01-23 01:40 [view ] -54.5892 141.2928 Aerosol Lab [view] Fault in2018_v01 Main output circuit breaker tripped.
IN2018_V01 2018-01-23 03:24 [view ] -54.9062 141.0923 Aerosol lab power [view] Atmospherics in2018_v01 Controlled shutdown for power replacements
IN2018_V02 2018-03-03 03:38 [view ] -43.2049 147.2844 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_01 (L1 and L2) START
IN2018_V02 2018-03-04 04:30 [view ] -46.4635 145.0233 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_01 (L1 and L2) STOP
IN2018_V02 2018-03-04 04:44 [view ] -46.4630 145.0230 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_02 (L1 and L2) START
IN2018_V02 2018-03-05 07:03 [view ] -46.9760 141.9620 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_02 (L1 and L2) STOP
IN2018_V02 2018-03-05 07:30 [view ] -46.9760 141.9620 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_01PB (L1 and L2), 3 min blank
IN2018_V02 2018-03-05 07:44 [view ] -46.9780 141.9620 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_03(L1 and L2)
IN2018_V02 2018-03-06 11:29 [view ] -47.0150 142.2410 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_03(L1 and L2) STOP
IN2018_V02 2018-03-06 11:50 [view ] -47.0120 142.2350 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_04(L1 and L2) START
IN2018_V02 2018-03-07 10:58 [view ] -46.9750 141.9320 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_04(L1 and L2) STOP
IN2018_V02 2018-03-07 11:22 [view ] -46.9760 141.9320 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_01EB (L1 and L2), ~24h exposure blank, START
IN2018_V02 2018-03-08 09:02 [view ] -46.4800 142.1120 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_01EB (L1 and L2), ~24h exposure blank, STOP
IN2018_V02 2018-03-08 09:26 [view ] -46.5230 142.0720 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_05 (L1 and L2), START
IN2018_V02 2018-03-09 06:40 [view ] -46.9920 141.9954 Aerosol Lab [view] Air Sampling in2018_v02 Ship exhaust sampling for TM, RV2018_V02_01SE START
IN2018_V02 2018-03-09 07:00 [view ] -46.9910 141.9930 Aerosol Lab [view] Air Sampling in2018_v02 Ship exhaust sampling for TM, RV2018_V02_01SE STOP
IN2018_V02 2018-03-09 07:10 [view ] -46.9902 141.9921 Aerosol Lab [view] Air Sampling in2018_v02 Ship exhaust sampling for TM, RV2018_V02_01SEPB, procedure blank in Aerosol Lab
IN2018_V02 2018-03-09 21:43 [view ] -46.0560 142.1810 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_05 (L1 and L2), STOP
IN2018_V02 2018-03-09 22:04 [view ] -46.0560 142.1810 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_06 (L1 and L2), START
IN2018_V02 2018-03-10 22:12 [view ] -45.5230 142.1590 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_06 (L1 and L2), STOP
IN2018_V02 2018-03-10 22:30 [view ] -45.5230 142.1590 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_02PB (L1 and L2), 3 minutes procedure blank
IN2018_V02 2018-03-12 01:55 [view ] -42.8864 147.3388 Aerosol Lab [view] Air Sampling in2018_v02 Ship exhaust sampling for TM, RV2018_V02_02SE START
IN2018_V02 2018-03-12 02:15 [view ] -42.8864 147.3388 Aerosol Lab [view] Air Sampling in2018_v02 Ship exhaust sampling for TM, RV2018_V02_02SE STOP
IN2018_V02 2018-03-12 05:05 [view ] -43.1180 147.4450 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_07 (L1 and L2), START
IN2018_V02 2018-03-12 06:53 [view ] -43.4200 147.4500 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_07 (L1 and L2), STOP
IN2018_V02 2018-03-12 07:08 [view ] -43.4630 147.4390 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_08 (L1 and L2), START
IN2018_V02 2018-03-13 21:41 [view ] -46.9950 142.0090 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_08 (L1 and L2), STOP
IN2018_V02 2018-03-13 22:13 [view ] -46.9970 142.0090 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_09 (L1 and L2), START
IN2018_V02 2018-03-14 02:38 [view ] -46.9850 141.9960 Aerosol Lab [view] Air Sampling in2018_v02 Ship exhaust sampling for TM, RV2018_V02_03SE START
IN2018_V02 2018-03-14 03:03 [view ] -46.9824 141.9966 Aerosol Lab [view] Air Sampling in2018_v02 Ship exhaust sampling for TM, RV2018_V02_03SE STOP
IN2018_V02 2018-03-14 21:26 [view ] -46.9940 142.0010 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_09 (L1 and L2), STOP
IN2018_V02 2018-03-14 21:40 [view ] -46.9940 142.0010 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_03PB (L1 and L2), 3minutes procedure blank
IN2018_V02 2018-03-14 21:57 [view ] -46.9940 142.0010 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_10 (L1 and L2), START
IN2018_V02 2018-03-16 11:25 [view ] -46.9930 141.8950 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_10 (L1 and L2), STOP
IN2018_V02 2018-03-16 11:45 [view ] -46.9840 141.8920 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_11 (L1 and L2), START
IN2018_V02 2018-03-17 12:31 [view ] -45.8710 141.7760 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_11 (L1 and L2), STOP
IN2018_V02 2018-03-17 12:48 [view ] -45.9140 141.7740 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_12 (L1 and L2), START
IN2018_V02 2018-03-20 01:45 [view ] -44.6950 143.2660 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_04PB (L1 and L2), 3minutes procedure blank
IN2018_V02 2018-03-20 01:47 [view ] -44.6950 143.2660 Aerosol Lab [view] Air Sampling in2018_v02 Aerosol sampling for TM, RV2018_V02_12 (L1 and L2), STOP
IN2018_V04 2018-09-21 18:41 [view ] -40.1233 153.4418 Aerosol Lab [view] Maintenance Atmospherics_in2018_v04 Velocity measurement on manifold. Plug removed for 20 minutes. Max velocity reading: 19.3m/s
IN2018_V05 2018-10-21 16:05 [view ] -54.9627 149.0070 Aerosol Lab [view] Observation Atmospherics_in2018_v05 Wrapped nafion dryers in insulation (partially) to see if it explains dewpoint cycling of sample air
IN2018_V05 2018-10-24 12:44 [view ] -55.6572 150.7615 Aerosol Lab [view] Maintenance Atmospherics_in2018_v05 emptied inlet water drain bottle
IN2018_V06 2018-11-26 22:13 [view ] -44.3050 147.1747 Aerosol Lab [view] Observation in2018_v06_Atmospherics lab check. all good
IN2018_V06 2018-11-28 19:59 [view ] -44.2985 147.2678 Aerosol Lab [view] Observation in2018_v06_Atmospherics Lab check. All good CPC butanol should be changed tomorrow
IN2018_V06 2018-12-07 03:27 [view ] -44.2317 146.1712 Aerosol Lab [view] Maintenance in2018_v06_Atmospherics Manifold catchpot emptied
IN2018_V06 2018-12-12 01:42 [view ] -44.2390 147.4658 Aerosol Lab [view] Maintenance in2018_v06_Atmospherics Installed MAAP bypass MFC Flow rate set to 1LPM. Serial number 179326
IN2018_V06 2018-12-18 18:58 [view ] -42.9947 147.3830 Aerosol Lab [view] Stop in2018_v06_Atmospherics shut down for christmas port period
IN2019_V01 2019-02-17 04:50 [view ] -66.3613 152.3722 Aerosol Dry Air Supply [view] Outage in2019_v01_Atmospherics main tap turned off for short period to install Tee to feed dry air to airchem lab
IN2019_V06 2019-10-28 12:34 [view ] -13.4892 121.5302 Aerosol Lab [view] Change in2019_v06_Atmospherics Air conditioning vent direction changed I changed the direction of this a few days ago so that it wasn& 39;t directly blowing onto the PTRMS line, since it was causing condensation. I think this meant it was blowing directly onto the MAAP and aerosol sample lines instead, which increased the RH in those lines to 60%. I think will came in about an hour or so ago and redirected one of them down again onto the PTRMS line, which has resulted in condensation again. I have changed it back for now. Now 12:42 - I have just added a bunch of towels over the insulated PTRMS line and pointed the AC fan down.
IN2019_V06 2019-10-30 10:34 [view ] -13.7980 121.9578 Aerosol Lab [view] in2019_v06_Atmospherics Wrapped the PTRMS line in bubble wrap and readjusted the AC vents to the correct downward direction again
IN2019_V06 2019-11-04 00:43 [view ] -14.2902 121.4613 Aerosol Lab [view] in2019_v06_Atmospherics Packing up AIRBOX O3 and NOx in aerosol lab
IN2019_V06 2019-11-04 00:47 [view ] -14.2902 121.4613 Aerosol Lab [view] in2019_v06_Atmospherics Removed AIRBOX O3 exhaust line from butanol exhaust and plugged it with paper after a small butanol smell was detected in that corner
IN2019_V06 2019-11-06 23:23 [view ] -13.4892 121.5302 Aerosol Lab [view] Change in2019_v06_Atmospherics Air conditioning vent direction changed slightly I discovered that the filters sitting under the AC were soaking wet this morning. Since there are no waves, its likely it was condensation. I have changed the direction slightly so its not directly on the cartridges. Not sure why this is the first time this has happened.
IN2019_V06 2019-11-13 19:01 [view ] -11.2115 133.4392 Aerosol Lab [view] Change in2019_v06_Atmospherics Insulation installed on nafions The sample line temperatures were fluctuating significantly due to the air conditioning. I have insulated the nafions as they are the largest surface. The temps are now stable/reflective of outside temp.
IN2019_V06 2019-11-20 03:43 [view ] -12.0930 130.6395 Aerosol Lab [view] Change in2019_v06_Atmospherics changes to air flow I have applied some makeshift
IN2019_V06 2019-12-17 00:03 [view ] Aerosol Lab [view] Change in2019_v06_Atmospherics Lab temp reduced to 22 for port period
IN2020_E01 2020-07-29 09:16 [view ] -42.9727 147.3807 Aerosol Lab [view] Change in2020_e01_Atmospherics Aerosol lab Interface unit - additional sensor installed Rotronics probe installed to monitor labratory temp hum and dew point
IN2020_V01 2020-01-13 01:01 [view ] -41.1913 104.4422 Aerosol Lab [view] Change in2020_v01_Atmospherics Changed Aerosol Lab temperature to 22 degrees Ian requested the lab temperature to be set to 22 degrees once the dew point outside is consistently below 20 degrees
IN2021_V01 2021-01-29 17:09 [view ] -44.4872 145.9657 Aerosol Lab [view] Change Atmospherics AC setpoint reduced to 22 degrees
IN2021_V01 2021-01-31 02:38 [view ] -48.5908 140.5403 Aerosol Lab [view] Change Atmospherics AC unit fan set to always on, temp set to 21 degrees for colder climate AC was getting to temp then automatically turning the fan off, I think this was resulting in a pocket of cooled air around the AC unit. The fan is now set to always on to ensure that air will continue circulating around the AC unit while the AC unit is not heating or cooling.
IN2021_V01 2021-03-22 18:50 [view ] -46.1260 144.6073 Aerosol Lab [view] Change Atmospherics Changed AC set point to 21 degrees Lab temp was very high (29 degrees) - set point was 24 degrees and I changed it to 21
IN2021_V03 2021-05-13 23:12 [view ] -27.4018 155.1400 Aerosol Lab [view] Change Atmospherics AC temp 25C
IN2021_V04 2021-07-19 14:38 [view ] -20.8954 108.1953 Aerosol Lab [view] Atmospherics LICOR connected on aerosol line in place of CCN
IN2021_V04 2021-07-26 07:48 [view ] -40.7068 132.2145 Aerosol Lab [view] Change Atmospherics Air con set to 22 degrees with fan always on (lab was getting too cold)