Survey | When (UTC) | Latitude | Longitude | Instrument | Identifier | Action | Log | Subject / Comments |
---|---|---|---|---|---|---|---|---|
IN2016_V03 | 2016-04-30 19:02 [view ] | -58.2661 | 163.1072 | radon detector [view] | restarted | in2016_v03 | Restart radon detector The radon detector tank pressures had dropped out of acceptable range for 24hrs so the instrument was rebooted. ANSTO have been notified. | |
IN2016_V03 | 2016-05-18 06:50 [view ] | -50.5003 | -170.0086 | radon detector [view] | Software Update | in2016_v03 | Installed fix for pressure drop - CR800_May16.CR8 | |
IN2016_V03 | 2016-05-28 13:55 [view ] | -42.3593 | -175.0147 | radon detector [view] | Fault | in2016_v03 | EXFlow-25 and Tankp-0. | |
IN2016_V03 | 2016-06-13 13:42 [view ] | -18.7463 | -170.0291 | radon detector [view] | Repair | in2016_v03 | Water Ingress Water found in sampling lines. lines opened to remove water. Inlet filter also had water/condensation build up. | |
IN2016_V03 | 2016-06-27 13:56 [view ] | -11.1771 | -178.0020 | radon detector [view] | maintenance | in2016_v03 | Water Trap Water trap emptied, sampling valve turned off for drain. | |
IN2016_V05 | 2016-09-30 08:14 [view ] | -20.9938 | 150.1988 | radon detector [view] | error | in2016_v05 | tanks pressure low | |
IN2017_T01 | 2017-09-28 02:04 [view ] | -19.0105 | 152.9404 | Radon detector [view] | maintenance | in2017_t01 | Increased tank pressure to 120 as per ANSTO instructions. | |
IN2017_T01 | 2017-10-08 01:32 [view ] | -17.9043 | 121.9987 | Radon detector [view] | calibration | in2017_t01 | Started Calibration, set background for 1600. | |
IN2017_V01 | 2017-03-04 06:35 [view ] | -44.4885 | 145.8333 | Radon detector [view] | calibration | in2017_v01 | Calibration started | |
IN2017_V02 | 2017-03-26 13:05 [view ] | -45.2842 | 144.1653 | Radon detector [view] | Stop | in2017_v02 | Program stopped through Marine Park | |
IN2017_V02 | 2017-03-26 19:00 [view ] | -44.7447 | 146.6547 | Radon detector [view] | Start | in2017_v02 | Restarted after leaving marine park | |
IN2017_V04 | 2017-08-31 09:02 [view ] | -26.7293 | 153.2333 | Radon detector [view] | calibration | in2017_v04 | Manual calibration - injection on Flush was started at 0400. Will turn off injection in 5 hours. | |
IN2017_V05 | 2017-10-12 00:20 [view ] | -19.2703 | 118.9804 | Radon detector [view] | Reboot | in2017_v05 | Radon detector PC was found to be disconnected (restarted) Plug for the power board had come out of the wall slightly. Plugged it back in and logged into the PC. Unknown how long it was down for. | |
IN2018_T02 | 2018-05-20 19:56 [view ] | -43.0107 | 147.3813 | Radon Detector [view] | Maintenance | Atmospherics_in2018_t02 | Started calibration run. Overlooked scheduling a Background - manually start this tomorrow | |
IN2018_V05 | 2018-10-16 11:00 [view ] | -43.9061 | 147.7681 | Radon Detector [view] | Maintenance | Atmospherics_in2018_v05 | calibration run Calibration was started about 6 hours after ship left port | |
IN2019_E01 | 2019-07-31 23:49 [view ] | -13.8880 | 146.0650 | Radon Detector [view] | Start | in2019_e01_Atmospherics | Started following dry dock | |
IN2019_E01 | 2019-07-31 23:50 [view ] | -13.8880 | 146.0650 | Radon Detector [view] | Maintenance | in2019_e01_Atmospherics | Calibration started Calibration started. Flush for 6 hours then cal for 6 hours. Note we do not normally flush, as the dry air is constantly flushiing the source. The flush was added in this instance as an extra precaution | |
IN2019_T03 | 2019-12-22 22:30 [view ] | -12.4718 | 130.8468 | Radon Detector [view] | Maintenance | in2019_t03_Atmospherics | Started Calibration | |
IN2019_V03 | 2019-05-15 11:50 [view ] | -34.7075 | 113.3077 | Radon Detector [view] | Maintenance | in2019_v03_Atmospherics | Calibration start Forgot to do this yesterday. Better late than never | |
IN2019_V03 | 2019-06-12 08:20 [view ] | -28.4595 | 113.4812 | Radon Detector [view] | Maintenance | in2019_v03_Atmospherics | Calibration started | |
IN2019_V05 | 2019-09-28 22:29 [view ] | -26.7250 | 153.2277 | Radon Detector [view] | Maintenance | in2019_v05_Atmospherics | Calibration start | |
IN2019_V06 | 2019-11-14 22:07 [view ] | -11.2073 | 133.4450 | Radon Detector [view] | Maintenance | in2019_v06_Atmospherics | Water trap drained small amount of very carbon filled water drained from forward trap. pump turned off briefly to facilitate this | |
IN2019_V06 | 2019-11-23 00:41 [view ] | -12.3125 | 130.4000 | Radon [view] | AIRBOX_in2019_v06 | Calibration and background details - 20/11/19 23:37: Ext and int flow cables unplugged (Manual background procedure started) - 22/11/19 00:02: Ext and int flow cables replugged (Manual background ended), pump switch turned on, starting manual calibration procedure 22/11/19 09:06: Valve switch turned on, continuing manual calibration procedure 22/11/19 23:09: Valve and pump switches off, ending manual calibration procedure - Following these, external flow has dropped to around 0-2 L/min, about 3 hrs before manual calib was stopped. Ext flow recovered successfully after manual background, but seems to have failed now. | ||
IN2020_E01 | 2020-07-29 05:47 [view ] | -42.9730 | 147.3801 | Radon Detector [view] | Maintenance | in2020_e01_Atmospherics | calibration started | |
IN2020_V01 | 2020-01-14 16:30 [view ] | -45.0975 | 99.2688 | Radon Detector [view] | Stop | in2020_v01_Atmospherics | Turned off to repair detector head Head fell off. | |
IN2020_V01 | 2020-01-17 15:45 [view ] | -53.5345 | 86.7512 | Radon Detector [view] | Start | in2020_v01_Atmospherics | Restarted after installing PMM. | |
IN2020_V01 | 2020-03-05 16:01 [view ] | -31.9742 | 115.2528 | Radon Detector [view] | Change | in2020_v01_Atmospherics | Cal started. Background scheduled for 0400 06/03 | |
IN2020_V06 | 2020-10-08 03:45 [view ] | -43.1007 | 147.4360 | Radon Detector [view] | Maintenance | in2020_v06_Atmospherics | Calibration Started | |
IN2020_V08 | 2020-12-04 15:18 [view ] | -44.0138 | 147.5518 | Radon Detector [view] | Maintenance | Atmospherics_in2020_v08 | calibration started | |
IN2020_V08 | 2021-01-04 05:37 [view ] | -57.9453 | 141.2662 | Radon Detector [view] | Fault | Atmospherics_in2020_v08 | Software "overrun" error. Restarted | |
IN2020_V09 | 2020-08-27 03:03 [view ] | -42.9513 | 147.3787 | Radon Detector [view] | Maintenance | in2020_v09_Atmospherics | Calibraton started Ian started the calibration from ashore | |
IN2020_V09 | 2020-09-02 03:50 [view ] | -46.8462 | 141.9123 | Radon Detector [view] | Outage | in2020_v09_Atmospherics | Stopped and started monitor software Stopped and started monitor software as values have not been updateing for a few hours | |
IN2020_V09 | 2020-09-02 05:30 [view ] | -46.8192 | 141.8655 | Radon Detector [view] | Start | in2020_v09_Atmospherics | outage resolved The fault was resolved. Looks like the lab door swung open hard in the storm, broke the door stop and dislodged the mains power for the radon detector unit. symptoms included no connection from the pc to the datalogger. Rod reconnected the power and it works now. | |
IN2020_V10 | 2020-11-21 20:53 [view ] | -42.8943 | 147.3532 | Radon Detector [view] | Maintenance | in2020_v10_Atmospherics | Radon calibration started | |
IN2021_T01 | 2021-06-11 16:38 [view ] | -24.5185 | 154.1685 | Radon Detector [view] | Maintenance | Atmospherics | calibration started | |
IN2021_T01 | 2021-06-15 04:24 [view ] | -16.6102 | 146.7145 | Radon Detector [view] | Fault | Atmospherics | Software crashed - restarted Overflow error | |
IN2021_T01 | 2021-06-18 10:19 [view ] | -10.6232 | 141.6373 | Radon Detector [view] | Start | Atmospherics | Restart after software run time error | |
IN2021_V01 | 2021-03-01 16:44 [view ] | -64.2567 | 70.0000 | Radon Detector [view] | Start | Atmospherics | Re-booted software The buffer overflow error is happening more frequently - so the PC was rebooted and the software restarted. | |
IN2021_V02 | 2021-04-14 18:04 [view ] | -43.3230 | 147.4857 | Radon Detector [view] | Start | Atmospherics | Restart after application showed overflow Restarted Radon application after software showed "overflow" error (no other info) | |
IN2021_V03 | 2021-05-08 05:19 [view ] | -43.2803 | 148.0370 | Radon Detector [view] | Fault | Atmospherics | Runtime error: software stopped. | |
IN2021_V03 | 2021-05-08 05:23 [view ] | -43.2810 | 148.0555 | Radon Detector [view] | Maintenance | Atmospherics | Calibration started | |
IN2021_V04 | 2021-07-20 22:44 [view ] | -26.3128 | 110.2978 | Radon Detector [view] | Maintenance | Atmospherics | Calibration started. | |
IN2021_V04 | 2021-07-20 22:45 [view ] | -26.3152 | 110.2983 | Radon Detector [view] | Maintenance | Atmospherics | Calibration started. | |
IN2021_V04 | 2021-07-21 13:02 [view ] | -28.5632 | 111.2110 | Radon Detector [view] | Maintenance | Atmospherics | Radon detector background started |