Please login if you have access to particular applications.
Search for events by voyage, equipment, action or within comments.
Event count for each voyage, equipment or action is shown in ().
Events ordered by start time:- Click on column header to sort. Use link on 'When' to see full details. Use the CSV button at the bottom to export data.
Survey | When (UTC) | Latitude | Longitude | Instrument | Identifier | Action | Log | Subject / Comments |
---|---|---|---|---|---|---|---|---|
IN2015_C02 | 2015-12-21 11:39 [view ] | -33.0452 | 114.9560 | Ozone Detector [view] | Repair | in2015_c02 | The Ozone detector stopped reporting data a couple of days ago. Inspection showed the instrument had suffered an electrical fault. The instrument was repaired this afternoon and put back into operation around 0900 UTC 21/12/2015 | |
IN2016_E01 | 2016-08-19 05:36 [view ] | -43.8096 | 150.5877 | Ozone Detector [view] | Start | in2016_e01 | Turned power on This had been running but had been turned off | |
IN2016_V03 | 2016-06-13 13:38 [view ] | -18.7553 | -170.0301 | Ozone Detector [view] | Repair | in2016_v03 | Ozone turned off removed from lab to electronics lab for repair. Water ingress was present into sampling lines, detector cells. Dismantling and cleaning of cells required. Instrument now in operation working correctly. | |
IN2016_V03 | 2016-06-15 21:34 [view ] | -14.6661 | -169.9987 | Ozone Detector [view] | maintenance | in2016_v03 | Added water trap to ozone detector inlets (from ~20:00) | |
IN2016_V03 | 2016-06-16 17:05 [view ] | -13.5045 | -170.0025 | Ozone Detector [view] | Repair | in2016_v03 | removed water from input lines; off 16:39, on 16:51 | |
IN2016_V03 | 2016-06-17 21:00 [view ] | -11.4960 | -169.9990 | Ozone Detector [view] | Repair | in2016_v03 | Modified water trap on inlet, and connected to outside air input | |
IN2016_V03 | 2016-06-18 17:55 [view ] | -9.9228 | -169.6297 | Ozone Detector [view] | Repair | in2016_v03 | remove water from lines, off 17:56 on 17:40 | |
IN2016_V03 | 2016-06-19 22:02 [view ] | -8.0006 | -168.6160 | Ozone Detector [view] | Repair | in2016_v03 | turn off to clear lines of water, off 21:47, on 21:48 | |
IN2016_V03 | 2016-06-21 01:22 [view ] | -6.0000 | -168.7513 | Ozone Detector [view] | maintenance | in2016_v03 | turn off to clear lines of water, off 00:03, on 00:06 | |
IN2016_V03 | 2016-06-21 21:24 [view ] | -4.5008 | -168.7501 | Ozone Detector [view] | in2016_v03 | turn off to clear lines of water, off 21:03, on 21:06 | ||
IN2016_V03 | 2016-06-22 21:55 [view ] | -2.5005 | -168.7499 | Ozone Detector [view] | Repair | in2016_v03 | turn off to clear lines of water, off 21:35, on 21:37 | |
IN2016_V03 | 2016-06-23 18:13 [view ] | -0.9995 | -168.7494 | Ozone Detector [view] | maintenance | in2016_v03 | turn off to clear lines of water, off 17:58, on 18:01 | |
IN2016_V03 | 2016-06-25 00:54 [view ] | -2.3356 | -170.6714 | Ozone Detector [view] | in2016_v03 | turn off to clear lines of water, off 00:43, on 00:47 also tightened some of the swageloks as pulled out when emptying traps. | ||
IN2016_V03 | 2016-06-25 13:47 [view ] | -4.2353 | -172.2364 | Ozone Detector [view] | in2016_v03 | turn off to clear lines of water | ||
IN2016_V03 | 2016-06-26 10:19 [view ] | -7.2030 | -174.6898 | Ozone Detector [view] | restarted | in2016_v03 | RVT1 & 2 restarted for removal of water build up in sampling line. | |
IN2016_V03 | 2016-06-27 00:00 [view ] | -9.1483 | -176.3064 | Ozone Detector [view] | in2016_v03 | turn off to clear lines of water, off 23:51, on 23:53 | ||
IN2016_V03 | 2016-06-27 13:55 [view ] | -11.1746 | -177.9999 | Ozone Detector [view] | maintenance | in2016_v03 | Empty water trap RVT1 and RVT2 off for water trap drain | |
IN2016_V03 | 2016-06-28 21:28 [view ] | -15.4690 | 178.2300 | Ozone Detector [view] | in2016_v03 | turn off to clear lines of water, on 20:41 needed to turn on only as was left off after last empty | ||
IN2017_T01 | 2017-09-30 16:04 [view ] | -10.1152 | 144.6446 | Ozone [view] | Shutdown | in2017_t01 | ozone water trap emptied, both monitors shutdown to empty trap to avoid water sucking through | |
IN2017_T01 | 2017-10-02 15:23 [view ] | -10.8129 | 136.9633 | Ozone [view] | outage | in2017_t01 | pump turned off to drain condensation | |
IN2017_T01 | 2017-10-03 21:47 [view ] | -11.1602 | 134.0769 | Ozone [view] | restarted | in2017_t01 | Ozone 1 no reading. Power Cycle and running again. | |
IN2017_T01 | 2017-10-06 05:34 [view ] | -13.3439 | 124.2218 | Ozone [view] | maintenance | in2017_t01 | Ozone 1, high readin. Empty water trap, power cycle and running again. | |
IN2018_T02 | 2018-05-20 19:50 [view ] | -43.0175 | 147.3825 | Ozone Monitor [view] | Stop | Atmospherics_in2018_t02 | End of voyage shutdown | |
IN2018_V05 | 2018-11-02 13:02 [view ] | -55.7838 | 153.6988 | Ozone Monitor [view] | Observation | Atmospherics_in2018_v05 | Consistently flagging low flow and pressure on both instruments | |
IN2019_T02 | 2019-10-14 00:42 [view ] | -12.4722 | 130.8460 | Ozone Monitor [view] | Maintenance | in2019_t02_Atmospherics | water trap empty, shut off pumps and inlet valve | |
IN2019_V03 | 2019-06-12 08:18 [view ] | -28.4563 | 113.4803 | Ozone Monitor [view] | Stop | in2019_v03_Atmospherics | Shut down for dry dock | |
IN2019_V06 | 2019-10-21 06:48 [view ] | -13.3242 | 123.7068 | Ozone Monitor [view] | Maintenance | in2019_v06_Atmospherics | pumps off, sample line tap off to drain condensation | |
IN2019_V06 | 2019-10-30 14:52 [view ] | -13.7985 | 121.9595 | Ozone Monitor [view] | Outage | in2019_v06_Atmospherics | inlet off inlet switched off to find issues with flow rate, no observed changes upstream of the Tee | |
IN2019_V06 | 2019-11-06 06:29 [view ] | -13.7985 | 121.9598 | Ozone Monitor [view] | Maintenance | in2019_v06_Atmospherics | Disconnected the lines momentarily on each before the filter to check flow issue. | |
IN2019_V06 | 2019-11-12 18:41 [view ] | -11.0712 | 133.3532 | Ozone Monitor [view] | Maintenance | in2019_v06_Atmospherics | Water trap emptied | |
IN2021_E03 | 2021-12-01 00:18 [view ] | -42.9757 | 147.3752 | Ozone Monitor [view] | Start | Atmospherics | RVT1 - restart | |
IN2021_T01 | 2021-06-16 23:42 [view ] | -11.8255 | 145.0505 | Ozone Monitor [view] | Fault | Atmospherics | air leak into the smps sample 16/6 23:26 - I just found that the air inlet tube was not connected properly. air has been sucked into the sampling system since I connected up the SMPS on 13/6 18:55. This will have affected the SMPS, and CPCs. It is possible that it will have affected the.nephelometer and CCNC. It is possible, but I think unlikely, that it also affected the ozone and greenhouse gas system, depending on the interactions in the manifold and sample lines. | |
IN2021_V02 | 2021-04-25 15:23 [view ] | -47.0442 | 142.3427 | Ozone Monitor [view] | Outage | Atmospherics | Power cycled because of low flow alarms | |
IN2021_V03 | 2021-05-15 03:45 [view ] | -27.0558 | 155.2445 | Ozone Monitor [view] | Maintenance | Atmospherics | Calibration OMCS test | |
IN2021_V03 | 2021-05-17 16:27 [view ] | -26.7685 | 154.8345 | Ozone Monitor [view] | Maintenance | Atmospherics | omcs testing zero | |
IN2021_V03 | 2021-05-31 22:41 [view ] | -27.2250 | 153.8650 | Ozone Monitor [view] | Stop | Atmospherics | zero's deactivated on RVT1 and RVT2 zeros deactivated-Zeros controlled from OMCS controller span on RVT2 continues | |
IN2021_V03 | 2021-06-02 20:25 [view ] | -27.2770 | 153.2530 | Ozone Monitor [view] | Change | Atmospherics | calibration | |
IN2021_V04 | 2021-07-12 12:00 [view ] | -13.7028 | 105.4995 | Ozone Monitor [view] | Outage | Atmospherics | RVT2 Turned off and disconnected to investigate flow fault | |
IN2021_V04 | 2021-07-12 14:57 [view ] | -13.7093 | 105.3438 | Ozone Monitor [view] | Outage | Atmospherics | Turned off RVT1 briefly | |
IN2021_V04 | 2021-07-13 02:18 [view ] | -11.9245 | 105.1877 | Ozone Monitor [view] | Maintenance | Atmospherics | Restarted 194 Board | |
IN2021_V04 | 2021-07-13 02:38 [view ] | -11.8608 | 105.1785 | Ozone Monitor [view] | Maintenance | Atmospherics | Ran pump efficiency test on 194 board RVT1 & 65533;pres 736.8 mm Hg, actual 736.8*sum 08bf RVT2 & 65533;pres 750.5 mm Hg, actual 750.5*sum 08b1 | |
IN2021_V04 | 2021-07-13 06:54 [view ] | -11.2367 | 104.8352 | Ozone Monitor [view] | Maintenance | Atmospherics | RVT2 pump repaired | |
IN2021_V04 | 2021-07-13 06:55 [view ] | -11.2357 | 104.8332 | Ozone Monitor [view] | Maintenance | Atmospherics | RVT2 "leak test and pump checkout" successful Did a leak test and pump checkout according to the manual page 157-158 | |
IN2021_V04 | 2021-07-13 07:24 [view ] | -11.2370 | 104.7712 | Ozone Monitor [view] | Maintenance | Atmospherics | Leak test and pump checkout on RVT1 and RVT2 after RVT2 pump replacement Did "leak test and pump checkout" according to pages 157-158 in the manual. ~50 seconds for RVT1 to reach 250 mmHg ~30 seconds for RVT2 to reach 250 mmHg | |
IN2021_V04 | 2021-07-13 07:55 [view ] | -11.2830 | 104.8228 | Ozone Monitor [view] | Maintenance | Atmospherics | Ran pump efficiency test on 194 board RVT1 - & 65533;pres 249.2 mm Hg, actual 249.2*sum 08b1 RVT2 - & 65533;pres 177.8 mm Hg, actual 177.8*sum 08bd | |
IN2021_V04 | 2021-07-25 05:01 [view ] | -38.5058 | 125.6330 | Ozone Monitor [view] | Stop | Atmospherics | RVT1 Turned off for maintanaince | |
IN2021_V04 | 2021-07-25 05:34 [view ] | -38.5517 | 125.7677 | Ozone Monitor [view] | Maintenance | Atmospherics | RVT1 Pump rebuilt New diaphragm and flapper plate. | |
IN2021_V04 | 2021-07-25 05:35 [view ] | -38.5533 | 125.7730 | Ozone Monitor [view] | Start | Atmospherics | RVT1 Turned on | |
IN2021_V04 | 2021-07-25 06:50 [view ] | -38.6597 | 126.0815 | Ozone Monitor [view] | Stop | Atmospherics | RVT1 turned off to check pump (not working after rebuild) | |
IN2021_V04 | 2021-07-25 17:00 [view ] | -39.4573 | 128.6133 | Ozone Monitor [view] | Start | Atmospherics | RVT1 reinstalled after pump repair | |
IN2021_V04 | 2021-07-25 17:13 [view ] | -39.4747 | 128.6675 | Ozone Monitor [view] | Observation | Atmospherics | Ran pump efficiency test on 194 board after RVT1 pump repair Ozone_pump_efficiency_result.JPG | |
IN2021_V04 | 2021-07-29 00:25 [view ] | -42.9328 | 147.3740 | Ozone Monitor [view] | Stop | Atmospherics | Both turned off. | |
IN2022_V02 | 2022-03-23 23:00 [view ] | Ozone monitor [view] | Ozone monitor event | |||||
IN2022_V02 | 2022-03-23 23:07 [view ] | Ozone monitor [view] | Ozone monitor event Changed filters on Ozone monitor 2. |