Event details
Voyage: | IN2021_V04 [details] |
Subject: | MFC is not working properly |
When: | 2021-07-04 16:07:43 UTC - entered into Elog/Everlog as 2021-07-04 16:07:43 |
Log: | in2021_v04/Atmospherics |
Equipment: | CPC 3750
Condensation Particle Counter (TSI 3750) [view] |
Position: |
Latitude: -11.68383 Longitude: 109.85617
- entered into Elog as 11 41.03 S, 109 51.37 E |
Action: | Fault |
Encoding: | HTML |
Steps taken: | |
Comments: | MFC was blocked with butanol and could not maintain the set flow rate. Cause could be a problem with catchpot drain system and is being investigated. Steps taken: - I cleaned the MFC and reinstalled - the flow rate appeared to be stable, however the CPC concentration was too high - Checked the flow rate with the Gilibrator and the actual flow rate was too high (~1440ccm) - I took the MFC apart again to ensure it was put together correctly, flow rate was still too high - I adjusted the set point of the MFC until the flow rate was 1000ccm (set point is now 0.66LPM) - did a zero check - |