Subject: | Beamformer started OK. Problem User |
When: | 2020-12-30 06:14:46 UTC - entered into Elog/Everlog as 2020-12-30 06:14:46 |
Log: | GSM_in2020_v08 |
Equipment: | ME70
Kongsberg ME70 shallow water multibeam scientific and fisheries echo sounder system [view] |
Position: |
Latitude: -55.84217 Longitude: 138.486
- entered into Elog as 55 50.53 S, 138 29.16 E |
Action: | Start |
Encoding: | HTML |
Comments: | Process- Start Beamformer software first... I think the problem was the user setting was my previous and it was in a pinging/recording state??? Use previous in2020_e01 (in2020_v08 should be good now) Open TRU On/Off Press START Should quickly go and configure the 25 cards and go green. Also power OFF and reseat RJ45 connectors may have helped?? Wait 30mins may help in restarting. |