Please login if you have access to particular applications.
Event details
Voyage: | IN2020_V08 [details] |
Subject: | Blocked Nozzle |
When: | 2020-12-21 11:05:23 UTC - entered into Elog/Everlog as 2020-12-21 11:05:23 |
Log: | Atmospherics_in2020_v08 |
Equipment: | CPC 3750
Condensation Particle Counter (TSI 3750) [view] |
Position: |
Latitude: -51.60067 Longitude: 138.493
- entered into Elog as 51 36.04 S, 138 29.58 E |
Action: | Fault |
Encoding: | HTML |
Comments: | CPC3750 is out of action due to what appears to be a blocked nozzle. Inlet pressure to the MFC has dropped to 750. A gradual decline is evident over 3 months, but most notably in the last month. Vacuum appears to have become sufficiently weak inside CPC to prevent butanol from being drawn in. I will inspect the system when possible. I have swapped over to the CPC3772 in the meantime. |