Subject: Flow Check on NanoSMPS
When: 2015-02-08 07:58:41 UTC

- entered into Elog/Everlog as 2/8/2015 7:58:41 AM
Log: IN2015_e01
Equipment: Other

Profiling Lagrangian Acoustic Optical System [view]
Station/Location: Aerosol lab
Position: Latitude: -64.5582 Longitude: 145.983 Position from underway logging
Action: Air Sampling
Event time:00am
7:00 Am checking long smps flows stoping SMPS Disconected SMPS line
restartting at 7:08 AM
Comments:
7:00 Am checking long smps flows stoping SMPS  Disconected SMPS lineShort SMPS restarted at  3461Long =SMPS restarted at  3399restartting at  7:08 AMThe CPC on the NanoSMPS after connecting back is showing counts all the time.Restarted the classifier and the CPC. It looks like the clasifier is arcing, large counts at voltages over 3000V. Will leave it to  1hour and restart afterwards. No scanning. 