Event details
Voyage: | IN2015_E01 [details] |
Subject: | Flow Check on NanoSMPS |
When: | 2015-02-08 07:58:41 UTC - entered into Elog/Everlog as 2/8/2015 7:58:41 AM |
Log: | IN2015_e01 |
Equipment: | Other
Profiling Lagrangian Acoustic Optical System [view] |
Station/Location: | Aerosol lab |
Position: | Latitude: -64.5582 Longitude: 145.983 Position from underway logging |
Action: | Air Sampling |
Event time: | 00am |
7: | 00 Am checking long smps flows stoping SMPS Disconected SMPS line |
restartting at 7: | 08 AM |
Comments: | 7:00 Am checking long smps flows stoping SMPS Disconected SMPS lineShort SMPS restarted at 3461Long =SMPS restarted at 3399restartting at 7:08 AMThe CPC on the NanoSMPS after connecting back is showing counts all the time.Restarted the classifier and the CPC. It looks like the clasifier is arcing, large counts at voltages over 3000V. Will leave it to 1hour and restart afterwards. No scanning. |