Please login if you have access to particular applications.
Event details
Voyage: | IN2016_V05 [details] | ||
Subject: | Reranged the vacuum lines. | ||
When: | 2016-10-07 01:50:40 UTC - entered into Elog/Everlog as 2016-10-07 01:30 01:45 | ||
Log: | in2016_v05 | ||
Equipment: | Voyage | ||
Station/Location: | Aerosol lab | ||
Position: | Latitude: -18.581 Longitude: 146.9732 Position from underway logging | ||
Action: | Air Sampling | ||
| |||
Comments: | Piccaro changed readings when we connected both the bypass flow (4.8lpm) and the dryer flow (4 lpm) to the clean line. We took of the drier flow of the clean and returned to the dirty vacuum line. Clean Vacum went down to -90kPa. Piccaro readings stabilised. |