Please login if you have access to particular applications.
Event details
Voyage: | IN2016_V05 [details] | ||
Subject: | Pump failure | ||
When: | 2016-10-06 20:50:48 UTC - entered into Elog/Everlog as 2016-10-06 20:50:48 | ||
Log: | in2016_v05 | ||
Equipment: | Piccaro
PICARRO G2301 Spectrometer [view] | ||
Position: | Latitude: -18.5825 Longitude: 146.9736 Position from underway logging | ||
Action: | restarted | ||
| |||
Comments: | The Picarro pump failed. From the Logs, probably at 0235 04/10/16 UTC. I have connected the picarro to the General Vacuum Line. Instrument is now running, cavity pressure locked at 140Torr, which is good. |