Event details

Voyage: IN2016_V03       [details]
When: 2016-06-24 02:38:14 UTC

- entered into Elog/Everlog as 2016-06-24 02:38:14
Log: in2016_v03
Equipment: Oxygen Analyser

Oxygen Analyser [view]
Position: Latitude: -0.414 Longitude: -168.749 Position from underway logging
Action: restarted
Comments:

Analyser found to be ina state with no FLSP flow and watchdog led lit. Restarted manifold box, and software, fault cleared and working okay now.

Trap also changed, and cleared small amoutn of condensation in inlet filters.