Please login if you have access to particular applications.
Event details
Voyage: | IN2019_V01 [details] | ||
Subject: | mouting location | ||
When: | 2019-01-31 11:34:17 UTC - entered into Elog/Everlog as 2019-01-31 11:34:17 | ||
Log: | in2019_v01_Atmospherics | ||
Equipment: | Barometer Aerosol | ||
Position: |
Latitude: -64.80383 Longitude: 143.4435
- entered into Elog as 64 48.23 S, 143 26.61 E | ||
Action: | Change | ||
| |||
Comments: | Barometer relocated back in aero-lab now 500mm lower than orignal position. The 14min artifact has now been removed, this was caused from the heating/cooling exchange units and or a.c unit. Barometer was previously mounted under an a.c vent, it is now furhter down in that position. To futher reduce the affect of temperature on the pressure transducer, a rudimentary enclosure was fashioned out of 100mm thick foam, protecting the unit from sharp changes in temperature - this worked well |