Event details
Voyage: | IN2019_V06 [details] |
Subject: | Bypass MFC swapped for upstream ssensor |
When: | 2019-11-15 20:21:27 UTC - entered into Elog/Everlog as 2019-11-15 20:21:27 |
Log: | in2019_v06_Atmospherics |
Equipment: | MAAP
Thermo Scientific Model 5012 MultiAngle Atmospheric Photometer [view] |
Position: |
Latitude: -11.2095 Longitude: 133.43867
- entered into Elog as 11 12.57 S, 133 26.32 E |
Action: | Change |
Parameters before change: | |
- Control: | 7.33 LPM Vol Flow |
Parameters after change: | |
Comments: | The MFC for the bypass ( 179327) had a sensor downstream of the valve. Parameters before change: - Control: 7.33 LPM Vol Flow - Vol Flow = 7.33 LPM - Mass FLow = 6.07 SLPM - pressure = 848mBarA I have swapped it for one with the sensor upstream ( 179324), making the measured volumetric flow more representative of VSCC flow (measured at same pressure). Parameters after change: - Control: 7.33 LPM Vol Flow - Vol Flow = 7.33 LPM - Mass FLow = 7.07 SLPM - Pressure = 987 mBarA Better for the VSCC |